Skip to main content

A rare case of persistent air leak: beware of all the tubes

Abstract

Background

Patients with acute respiratory failure, impaired consciousness, and impaired airway reflexes will require invasive mechanical ventilation. Monitoring of such patients is important. The use of ventilator scalars and loops help in monitoring, diagnosing the abnormality, and treating the patients effectively. We report a rare cause one should suspect in a case of persistent and fixed air leak in a patient requiring mechanical ventilation.

Case presentation

We describe a 28-year-old young patient requiring ventilator support due to neuromuscular weakness. His neuromuscular weakness was rapidly progressing involving the respiratory muscles. The patient was intubated and put on mechanical ventilator support. He was transferred from another health care center to our hospital. On evaluation, the patient was intubated with ETT no 8. The patient had persistent air leak as observed on the ventilator graphics. We checked for ETT cuff malfunction, ventilator circuit, catheter mount, and HME for any disconnection causing the leak. The air leak which we observed in our patient was due to the malpositioned Ryle’s tube.

Conclusions

Vigilant monitoring of patients requiring mechanical ventilation is necessary. For the evaluation of the cause of air leak, algorithmic approach will help in correctly identifying the abnormality.

Background

Monitoring of patients requiring mechanical ventilation is important. The use of Ventilator scalars and loops help in monitoring, diagnosing the abnormality, and treating the patients effectively. The use of Scalars helps in identifying abnormalities of resistance and compliance, presence of secretions, autopeep, and air leak. Similar changes will be observed in loops depending on the abnormality.

Presence of air leak is one of the common abnormalities one can come across. Possible causes of air leak could be ETT (Endotracheal tube) cuff leak, ventilator circuit traps which are loose, or ETT, which is of small size than the glottis. These abnormalities can be easily identified and can be corrected.

We report a rare case of persistent air leak in a 28-year-old patient (IEC ref no 006/2020).

Case presentation

A 28-year-old young male presented with the history of weakness in both lower limbs after viral illness. Neurological assessment showed weakness of both lower limbs which gradually progressed to involve the upper limbs. The possibility of ascending flaccid paralysis was considered. He received 3 cycles of plasmapheresis. Subsequently, the patient developed breathing difficulty and required ventilator support. He was transferred to our hospital. On evaluation, the patient had ET tube and Ryle’s tube and was connected to ventilator support.

Ventilator settings were as follows: PRVC mode Fio2 40%, RR 22, and TV 380 ml. On the ventilator graphics, air leak was noted (Fig. 1). ET tube position and cuff was checked. It was observed that the ET tube, no. 8, was fixed at around 18 cm. The clinical decision of check laryngoscopy after sedation and paralysis was made. On laryngoscopy, the ET tube cuff was above the glottis, and after deflation of cuff, it was difficult to advance the tube. So, the tube was changed to 8.5 no ETT and fixed at 22 cm H2O. Tube position was confirmed by auscultation and subsequently by ETCo2.

Fig. 1
figure 1

PRVC mode with air leak

After the change of ET tube, the leak persisted. So, the ventilator circuit and water traps were checked for any air leak. There was no subcutaneous emphysema, and lung ultrasound showed good lung sliding. So, we planned on doing CXR.

CXR showed that Ryle’s tube is malpositioned (Fig. 2). It was entering into the right main bronchus. Ryle’s tube’s distal end was open, and that was causing persistent leak. So, we planned to remove Ryle’s tube. After Ryle’s tube removal, the leak disappeared (Figs. 3 and 4).

Fig. 2
figure 2

CXR showing malpositioned Ryle’s tube

Fig. 3
figure 3

CXR after correct placement of Ryle’s tube

Fig. 4
figure 4

PRVC mode,  no air leak after the removal of Ryle’s tube

Discussion

This case represents various important lessons. First, one should follow algorithmic approach when evaluating for the cause of air leak (Fig. 5). The assessment should begin either from the patient end or ventilator end, so that all possible causes can be easily identified. From the patient end, when we assess, we look for the endotracheal or tracheostomy cuff for cuff leak. Cuff pressure can be checked with Lange’s pressure monitor. It can become a part of routine monitoring of patients. The catheter mount which has a flip valve can be the commonest site for air leak. The HME (heat moisture exchanger) filter which has a CO2 sampling port, if open, can cause air leak. Then, check the ventilator circuit and water traps. The attachment of ventilator circuit to the ventilator can be loose and may get disconnected. With the help of imaging such as CXR, ultrasound, or CT scan, one can identify the anatomical and parenchymal lung diseases causing air leak. If the patient has ICD (intercostal drain), if it is not positioned properly, one of the ports can be in the subcutaneous plane and can lead to air leak. In the patient of chest trauma, the presence of open chest wound can lead to inadequate minute ventilation due to air leak.

Fig. 5
figure 5

Algorithmic approach when evaluating for the cause of air leak

Regarding this particular case, malpositioning of Ryle’s tube was the cause of air leak. It is not a routine practice across all the hospitals to confirm Ryle’s tube position by CXR. The confirmation by auscultation with insufflation of air method is not specific to confirm the placement of tube as the test may be false positive, and also, it may not ensure the correct positioning of Ryle’s tube [1].

In the evaluation of air leak in patients requiring mechanical ventilation as shown by Gonzalez et al., air leak can cause persistent hypercapnia, and one can take practical measures to reduce the volume of air leak [2]. McInnis et al. described, a rare case of air leak due to esophageal malignancy invading the left main bronchus [3]. Lazarus et al. described the bronchoscopic management for persistent air leak [4]. There are various causes for persistent air leak in post thoracic surgery patients. Cerfolio’s classification is used for the grading of air leaks. Dugan et al. explained various methods used in the management of such patients [5].

Identification of correct etiology is important, in order to decide about the further plan of management. In our case, it was a malpositioned Ryle’s tube which led to air leak. Doing CXR for confirmation of tubes and catheters is required. Especially, in case of Ryle's tube, as auscultation is not enough to confirm the position, CXR is considered as an optimum and feasible method [1]. The article by Fan et al. describes various methods for the confirmation of Ryle’s tube position and advantages and disadvantages of each [1].

Conclusion

The cause of air leak in a patient described above was due to malpositioned Ryle’s tube. The physician should follow an algorithmic approach to identify the etiology correctly and manage accordingly. This case report also highlights the limitations of routine auscultation method used for confirming Ryle’s tube position.

Availability of data and materials

Available upon reasonable request

References

  1. Fan EM, Tan SB, Ang SY. Nasogastric tube placement confirmation: where we are and where we should be heading. Proc Singapore Healthcare. 2017;26(3):189–95. https://0-doi-org.brum.beds.ac.uk/10.1177/2010105817705141.

    Article  Google Scholar 

  2. Gonzalez J, Sharshar T, Hart N, Chadda K, Raphaël JC, Lofaso F. Air leaks during mechanical ventilation as a cause of persistent hypercapnia in neuromuscular disorders. Intensive Care Med. 2003;29(4):596–602. https://0-doi-org.brum.beds.ac.uk/10.1007/s00134-003-1659-5.

    Article  PubMed  Google Scholar 

  3. McInnis I, Walter R. 1165: an unusual case of air leak in a mechanically ventilated patient. Crit Care Med. 2019;47(1):559. https://0-doi-org.brum.beds.ac.uk/10.1097/01.ccm.0000551910.05745.5e.

    Article  Google Scholar 

  4. Lazarus DR, Casal RF. Persistent air leaks: a review with an emphasis on bronchoscopic management. J Thoracic Dis. 2017;9(11):4660–70. https://0-doi-org.brum.beds.ac.uk/10.21037/jtd.2017.10.122.

    Article  Google Scholar 

  5. Dugan KC, Laxmanan B, Murgu S, Hogarth DK. Management of persistent air leaks. Chest. 2017;152(2):417–23. https://0-doi-org.brum.beds.ac.uk/10.1016/j.chest.2017.02.020.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The authors would like to thank the MICU staff who helped in the data collection.

Funding

Not applicable

Author information

Authors and Affiliations

Authors

Contributions

AH has helped in collecting the data and writing the manuscript and approved the final draft. AF has helped in the data collection. AC has helped in the data collection. The author(s) read and approved the final manuscript.

Author’s information

AH is working as an associate  professor in the Department of Critical Care Medicine.

AF and AC were  working as senior residents in the department when study was done. 

Corresponding author

Correspondence to Amarja Ashok Havaldar.

Ethics declarations

Ethics approval and consent to participate

Formal consent was obtained from a legally acceptable representative (LAR), provisionally approved by the Institutional Ethics Committee.

Consent for publication

Written informed consent was obtained from the LAR for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Havaldar, A.A., Fazal, A.R. & Chennabasappa, A. A rare case of persistent air leak: beware of all the tubes. Int J Emerg Med 14, 58 (2021). https://0-doi-org.brum.beds.ac.uk/10.1186/s12245-021-00381-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12245-021-00381-6

Keywords