Chest tube placement in trauma patients: please use sonography

Case presentation

A 27-year-old patient with no known medical history was admitted for polytrauma resulting from a traffic accident. The initial thoraco-abdominal computed tomography (CT) scan revealed a minimal left pneumothorax, liver lacerations with intracapsular hematoma, and moderate hemoperitoneum. After 48 h, the patient’s respiratory status worsened. A subsequent chest CT scan identified a medium-sized right hemothorax leading to the insertion of a 28-Fr chest tube (Fig. 1, panel A). Approximately 400 mL of blood flowed freely through the chest tube before ceasing spontaneously. However, the chest tube was found to be dangerously low in its placement.

Fig. 1
figure 1

Illustrations of the insertion site (panel A), chest X-ray (panel B), and abdominal CT scan (panels C and D) depicting an intrahepatic aberrant chest tube highlighted by a black arrow in the chest X-ray

Diagnosis

A follow-up chest X-ray confirmed the misplaced chest tube, located intraabdominally (Fig. 1, panel B). Subsequent abdominal CT imaging confirmed the tube’s intrahepatic placement (Fig. 1, panels C and D). The blood flow through the intrahepatic tube likely corresponded to the intracapsular hematoma. The malpositioned tube was carefully removed under strict monitoring, and another chest tube was inserted under ultrasound (US) guidance, with the surgical team prepared for intervention if needed. The patient’s recovery was uneventful, and he was discharged after a few days.

This case underscores the value of using US guidance to prevent chest tube misplacement and emphasizes that relying solely on anatomical landmarks can lead to errors. In fact, ultrasound guidance allows the choice of the best puncture site and provides real-time visualization of the chest structures (Fig. 2, panels 1 and 2) [1]. Ultrasound guidance for chest drainage should be incorporated into routine medical practice [2]. However, comparative studies are needed to determine whether these approaches should be established as standard care.

Fig. 2
figure 2

1 Illustration depicting the recommended probe position during chest ultrasonography for pleural effusion drainage. Either a phased array or curved probe can be utilized; in this instance, a phased array probe was employed. 2 Chest ultrasound (intercostal view) displaying a substantial pleural effusion alongside a consolidated lung

This case highlights the importance of using US guidance to prevent chest tube misplacement, emphasizing that reliance solely on anatomical landmarks can lead to errors. Ultrasound guidance enables the selection of the best puncture site and provides real-time visualization of chest structures (Fig. 2, panels 1 and 2) [1]. Incorporating ultrasound guidance for chest drainage into routine medical practice is recommended. However, further comparative studies are necessary to determine whether these approaches should be established as standard care [2].

Availability of data and materials

Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.

References

  1. Peris A, Tutino L, Cianchi G, Gensini G. Ultrasound guidance for pleural-catheter placement. N Engl J Med. 2018;378:e19–e19.

    Article  PubMed  Google Scholar 

  2. Menegozzo CAM, Utiyama EM. Steering the wheel towards the standard of care: proposal of a step-by-step ultrasound-guided emergency chest tube drainage and literature review. Int J Surg. 2018;56:315–9.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

None

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

YA and FS wrote the manuscript text and prepared the images. All authors: YA, FS, ABo and ABe read and approved the final manuscript.

Corresponding author

Correspondence to Younes Aissaoui.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Written consent for publication was obtained from the patient.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Saroukh, F., Bouchama, A., Belhadj, A. et al. Chest tube placement in trauma patients: please use sonography. Int J Emerg Med 17, 20 (2024). https://0-doi-org.brum.beds.ac.uk/10.1186/s12245-024-00596-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12245-024-00596-3